Features:
● Self-diagnostic function:
(1) Core tests; (2) Automatic tests; (3) Extended tests; (4) Basic tests;
Monitor the status of GC continuously. Once finding faults, it will display the information, show the wrong zone and the settling method.
● Self-protective function:
(1) Overrun temperature protection; (2) Short circuit hint; (3) TCD filament protection; (4) FID flameout hint;
(5) PFD exposed-light protection; (6) Keyboard locked function utilizing password; more kinds of protect function ensured GC's normally running.
● Simple operation, powerful automation:
(1) All parameters can be entered through keyboard with prompt function;
(2) Storing 4 complete chromatography analysis methods which can be activated automatically;
(3) Can be connected with Autosampler;
(4) Parameters can be modified instantly while the GC is running;
(5) Chromatography analysis method can be activated for 99 times repeatedly at one time. It's especially suitable for in nobody guarding situation.
● More choices of injectors:
(1) On-Column Injector for Packed Column; (2) Flash Vaporization Injector for Packed Column
(3) Automatic or manual gas inject valve; (4) Headspace sampler; (5) Thermal desorption system
(6) Split/Splitless Capillary Injector; Three Injectors or two Split/Splitless Capillary Injectors can be fixed on GC.
● More choices of detectors:
(1) TCD; (2) FID; (3) ECD; (4)FPD; (5) TSD
Maximum of two TCD installed or three different kind of detectors.
● Reactor: (1) Internal; (2) External
● Time programming of Detectors:
Each of detectors has 5-ramp programmable time control, output-signal, attenuation range, and polarity can be changed automatically.
● Time programming of external events:
Provide 4 external events with 20-ramp programmable time control; the optional GC relays may be used to automate valves, to operate split/splitless capillary injectors, or drive auxiliary functions, or switch signal between detectors A and detector B during a run.
Technical Specification:
|
Model No: GC-002 |
|
|
|
Column Oven |
Operating temperature range |
ambient temperature to 420 ºC |
|
Owning dual opening-door at the back of the colume oven |
||
|
Structure |
Double – Pistin pump. |
|
|
Cooling time |
250 ºC to 50 ºC in 5 min |
|
|
Programmable oven with 4-ramp temperature control |
||
|
Temperature programming rate |
0.1~50 ºC /min |
|
|
Injector system |
Operating temperature range: ambient temperature to 400 ºC |
|
|
Thermal conductivity detector-TCD |
Maximum temperature |
400 ºC |
|
Sensitivity |
≥10000mv ●mL/mg ( for butane) |
|
|
Linear range |
105 |
|
|
Current range |
50 to 400mA |
|
|
Filament protection |
If He or H2 carrier gas flow in the detector cell is off for 4 minutes, filament power automatically shuts off. |
|
|
Flame ionization detector-FID |
Maximum temperature |
400 ºC |
|
Minimum detectivity |
≤5×10-12g/s (n-C16) |
|
|
Linear range |
107 |
|
|
Flameout hint: monitoring the status of GC continuously, after being turned on. |
||
|
Electron capture detector-ECD |
Maximum temperature |
400 ºC |
|
Minimum detectivity |
≤0.1pg/mL (γ-666) |
|
|
Linear range |
104 |
|
|
Emission source |
11mC63Ni |
|
|
Flame photometric detector-FPD Column Oven |
Maximum temperature |
400 ºC |
|
Minimum detectivity |
[P]≤2×10-12g/s (tributyl phosphate), [S]≤2×10-10g/s (p-methylsulfurphosoharous) |
|
|
Linear range |
[P]: 105 [S]: 103 |
|
|
Two air-hydrogen flames |
For large volume injection or trace sample injection analyses; |
|
|
Single air-hydrogen flame |
sulfur sensitivity can be increased; |
|
|
Direct output: for phosphorus or sulfur analyses; |
||
|
Square-root output |
for sulfur analyses; |
|
|
Exposed-light protection |
when exposed-light current≥ 12μA, high voltage will be cut off automatically. |
|
|
Nitrogen-phosphorus detector-TSD |
Maximum temperature |
400 ºC |
|
Minimum detectivity |
[N]≤2×10-13g/s (azobenzene), [P]≤1×10-13g/s (malathion) |
|
|
Linear range |
[N]: 105 [P]: 104 |
|