Single Beam Visible Spectrophotometer LVS-102

Labocon Single Beam Visible spectrophotometers LVS-100 Series offers measurement range of 320-1100 nm at bandwidth of 4nm; which provides high accuracy, reproducibility and high throughput at each run.

These LVS-100 Series are inbuilt with highly durable tungsten halogen lamp, which best suitable for:

• For measurement of DNA, RNA and oligonucleotide and Protein

• Proteins can be quantitated by BCA, Biuret, Bradford, Direct UV and Lowry

• The OD of bacterial cell culture solutions can be measured to optimize harvest times

• For the determination of metal and organic non-metal analytes in water

• Ideal to be use in petrochemical labs as they are solvent resistant such as Naphthalene in aviation

Wavelength Range 320-1100 nm
Wavelength Accuracy ±2 nm
Wavelength Repeatability ≤1 nm
Spectral Bandwidth 4 nm
Stability ±0.002 A/hr at 500 nm, 0 A
Baseline Flatness None
Photometric Range -0.3 to 3 A; 0-200 %T; -9999 to 9999 C
Photometric Accuracy ±0.5 %T (0-100%T)
Photometric Repeatability ±0.2 %T (0-100%T)
Photometric Mode None
Monochromator Littrow Diffraction Monochromator
Stray Light ≤0.2 %T (220nm, 340nm)
Light Source -
Display LCD
Overall Dimension 470x380x200 mm
Power -
Power Supply 240V 60Hz
Weight 14 kg
Catalog No. 9141146164

Labocon Visible Spectrophotometer LVS-100 Series is widely used in educational institutes, medical and health clinics, biochemistry field, chemical engineering, environmental protection, food industry, pharmaceutical industry, quality control departments, etc.

• Wavelength range

  1. 320nm-1100nm

• Optical System

  1. High quality Czerny Turner Diffraction Gratingmonochromator ensures high resolution, high photometric accuracy and low stray light

• Bandwidth

  1. 4nm

• Detector

  1. Silicon Photodiode

• Measurement Range with ultra-low Stray Light

  1. With its ultra-low stray light (≤0.2 %T (220nm, 340nm)) is achieved on wavelength range of 320 - 1100 nm

• Sample size

  1. A wide range of standard cuvettes can be used based on the sensitivity or sample volume requirements.

• Sophisticated and simplified sample analysis software

  1. The Sample analysis software deliver scanning, fixed wavelength analysis, quantitative analysis, data collection, storage, export, and reporting.
  2. With additional feature of an easy access USB port available in the unit, which enables results to be stored directly to a USB memory stick for easy transfer of data.

• Manual wavelength setting

• Microprocessor controlled technology

• Can save 200 standard curve and 500 groups of data

• USB data port

• Built-in printer and anti-controlled PC software